Digital MultiMeter  
Model No. 82140  
and  
Voltage Detector  
Model No. 82174  
Safety  
CAUTION: Read, understand and  
follow Safety Rules and Operating  
Instructions in this manual before  
using this product.  
Operation  
Maintenance  
Español  
© Sears, Roebuck and Co., Hoffman Estates, IL 60179 U.S.A.  
1
 
ONE YEAR FULL WARRANTY  
ONE YEAR FULL WARRANTY ON CRAFTSMAN MANUAL RANGING  
MULTIMETER  
If this CRAFTSMAN Manual Ranging MultiMeter fails to give complete  
satisfaction within one year from the date of purchase, RETURN IT TO  
THE NEAREST SEARS STORE OR OTHER CRAFTSMAN OUTLET IN  
THE UNITED STATES, and Sears will replace it, free of charge.  
If this CRAFTSMAN Manual Ranging MultiMeter is used for commercial  
or rental purposes, this warranty applies for 90 days from the date of  
purchase.  
This warranty gives you specific legal rights, and you may also have  
other rights which vary from state to state  
Sears, Roebuck and Co., Dept. 817WA, Hoffman Estates, IL 60179  
For Customer Assistance Call 9am-5 PM (EST)  
Monday through Friday 1-888-326-1006  
WARNING: USE EXTREME CAUTION IN THE USE OF THIS DEVICE.  
Improper use of this device can result in injury or death. Follow all  
safeguards suggested in this manual. In addition to the normal safety  
precautions used in working with electrical circuits. DO NOT service this  
device if you are not qualified to do so.  
3
 
SAFETY INSTRUCTIONS  
This meter has been designed for safe use, but must be operated with  
caution. The rules listed below must be carefully followed for safe operation.  
1. NEVER apply voltage or current to the meter that exceeds the  
specified maximum:  
Input Limits  
Function  
Maximum Input  
VAC  
600V DC/AC  
V DC or V AC  
mA DC  
600V DC/AC, 200Vrms on 200mV range  
200mA 250V fast acting fuse  
A DC  
10A 250V fast acting fuse(30 seconds max  
every 15 minutes)  
Resistance, Continuity 250Vrms for 15sec max  
2. USE EXTREME CAUTION when working with high voltages.  
3. DO NOT measure voltage if the voltage on the "COM" input jack  
exceeds 500V above earth ground.  
4. NEVER connect the meter leads across a voltage source while the  
function switch is in the current, resistance, or diode mode. Doing so  
can damage the meter.  
5. ALWAYS discharge filter capacitors in power supplies and disconnect  
the power when making resistance or diode tests.  
6. ALWAYS turn off the power and disconnect the test leads before  
opening the doors to replace the fuse or batteries.  
7. NEVER operate the meter unless the back cover and the battery and  
fuse doors are in place and fastened securely.  
4
 
SAFETY SYMBOLS  
This symbol adjacent to another symbol, terminal or  
operating device indicates that the operator must refer  
to an explanation in the Operating Instructions to avoid  
personal injury or damage to the meter.  
This WARNING symbol indicates a potentially  
hazardous situation, which if not avoided, could result  
in death or serious injury.  
WARNING  
CAUTION  
This CAUTION symbol indicates a potentially  
hazardous situation, which if not avoided, may result  
damage to the product.  
MAX  
500V  
This symbol advises the user that the terminal(s) so  
marked must not be connected to a circuit point at  
which the voltage with respect to earth ground  
exceeds (in this case) 500 VAC or VDC.  
This symbol adjacent to one or more terminals  
identifies them as being associated with ranges that  
may, in normal use, be subjected to particularly  
hazardous voltages. For maximum safety, the meter  
and its test leads should not be handled when these  
terminals are energized.  
This symbol indicates that a device is protected  
throughout by double insulation or reinforced insulation.  
5
 
CONTROLS AND JACKS  
1. LCD Display  
2. Function switch  
3. COM jack  
1
2
4. 10A jack  
5. Positive jack  
Note: Tilt stand, fuse and battery  
compartment are on rear of unit.  
3
4
5
SYMBOLS AND ANNUNCIATORS  
•)))  
Continuity  
Diode test  
micro (amps)  
milli ( volts, amps)  
kilo (ohms)  
m
k
ꢂꢂꢃꢂ  
ohms  
VDC  
VAC  
ADC  
BAT  
volts direct current  
volts alternating current  
amps direct current  
Battery test  
6
 
SPECIFICATIONS  
Function  
DC Voltage  
(V DC)  
Range  
Resolution  
0.1mV  
1mV  
0.01V  
0.1V  
Accuracy  
(0.5% reading + 2 digits)  
200mV  
2000mV  
20V  
200V  
600V  
1V  
AC Voltage  
(V AC)  
200V  
600V  
0.1V  
1V  
(1.2% reading + 10 digits  
(50/60Hz)  
DC Current  
(A DC)  
2000A  
20mA  
200mA  
10A  
200ꢃ  
2000ꢃ  
20kꢃ  
200kꢃ  
2000kꢃ  
9V  
1A  
(1.0% reading + 2 digits)  
10A  
100A  
10mA  
0.1ꢃ  
1ꢃ  
0.01kꢃ  
0.1kꢃ  
1kꢃ  
(1.2% reading + 2 digits)  
(2.0% reading + 2 digits)  
(0.8% reading + 2 digits)  
Resistance  
Battery Test  
(1.0% reading + 2 digits)  
(1.0% reading + 2 digits)  
10mV  
1.5V  
10mV  
NOTE: Accuracy specifications consist of two elements:  
(% reading) – This is the accuracy of the measurement circuit.  
(+ digits) – This is the accuracy of the analog to digital converter.  
NOTE: Accuracy is stated at 65oF to 83oF (18oC to 28oC) and less than  
75% RH.  
7
 
SPECIFICATIONS  
Diode Test  
Test current of 1mA maximum, open circuit  
voltage 2.8V DC typical  
Continuity Check  
Audible signal will sound if the resistance is  
less than approximately 30  
9V (6mA); 1.5V (100mA)  
>1M  
45Hz to 450Hz  
Battery Test current  
Input Impedance  
ACV Bandwidth  
DCA voltage drop  
Display  
200mV  
3 ½ digit, 2000 count LCD, 0.5” digits  
“1” is displayed  
Automatic (no indication for positive polarity);  
Minus (-) sign for negative polarity.  
2 times per second, nominal  
“BAT” is displayed if battery voltage drops  
below operating voltage  
Overrange indication  
Polarity  
Measurement Rate  
Low Battery Indication  
Battery  
Fuses  
one 9 volt (NEDA 1604) battery  
mA, µA ranges; 0.2A/250V fast blow  
A range; 10A/250V fast blow  
41oF to 104oF (5oC to 40oC)  
-4oF to 140oF (-20oC to 60oC)  
Max 80% up to 87ºF (31ºC) decreasing linearly  
to 50% at 104ºF (40ºC)  
Operating Temperature  
Storage Temperature  
Operating Humidity  
Storage Humidity  
Operating Altitude  
Weight  
<80%  
7000ft. (2000) meters maximum.  
9.17 oz. (260g).  
Size  
4.78” x 2.38” x 1.57” (121.5mm x 60.6mm x  
40mm)  
Safety  
For indoor use and in accordance with  
Overvoltage Category II, Pollution Degree 2.  
Category II includes local level, appliance,  
portable equipment, etc., with transient  
overvoltages less than Overvoltage Category III.  
8
 
BATTERY INSTALLATION  
WARNING: To avoid electric shock, disconnect the test leads from any  
source of voltage before removing the battery door.  
1. Disconnect the test leads from the meter.  
2. Remove the protective rubber holster (if installed).  
3. Open the battery door by loosening the screw using a  
Phillips head screwdriver.  
4. Insert the battery into battery holder, observing the  
correct polarity.  
5. Put the battery door back in place. Secure with the  
screw.  
WARNING: To avoid electric shock, do not operate the meter until the  
battery door is in place and fastened securely.  
NOTE: If your meter does not work properly, check the fuses and batteries  
to make sure that they are still good and that they are properly inserted.  
9
 
OPERATING INSTRUCTIONS  
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC,  
are very dangerous and should be measured with great care.  
1. ALWAYS turn the function switch to the OFF position when the meter is  
not in use.  
2. If “OL” appears in the display during a measurement, the value exceeds  
the range you have selected. Change to a higher range.  
NOTE: On some low AC and DC voltage ranges, with the test leads not  
connected to a device, the display may show a random, changing reading.  
This is normal and is caused by the high-input sensitivity. The reading will  
stabilize and give a proper measurement when connected to a circuit.  
DC VOLTAGE MEASUREMENTS  
CAUTION: Do not measure DC voltages if a motor on the circuit is  
being switched ON or OFF. Large voltage surges may occur that can  
damage the meter.  
1. Set the function switch to the highest V DC  
position.  
2. Insert the black test lead banana plug into the  
negative (COM) jack.  
Insert the red test lead banana plug into the  
positive (V) jack.  
3. Touch the black test probe tip to the negative side  
of the circuit.  
Touch the red test probe tip to the positive side of  
the circuit.  
4. Read the voltage in the display. Reset the  
function switch to successively lower V DC  
positions to obtain a higher resolution reading.  
The display will indicate the proper decimal point and value. If the  
polarity is reversed, the display will show (-) minus before the value.  
10  
 
AC VOLTAGE MEASUREMENTS  
WARNING: Risk of Electrocution. The probe tips may not be long  
enough to contact the live parts inside some 240V outlets for  
appliances because the contacts are recessed deep in the outlets. As  
a result, the reading may show 0 volts when the outlet actually has  
voltage on it. Make sure the probe tips are touching the metal contacts  
inside the outlet before assuming that no voltage is present.  
CAUTION: Do not measure AC voltages if a motor on the circuit is  
being switched ON or OFF. Large voltage surges may occur that can  
damage the meter.  
1. Set the function switch to the highest V AC position.  
2. Insert the black test lead banana plug into the  
negative (COM) jack.  
Insert red test lead banana plug into the positive (V)  
jack.  
3. Touch the black test probe tip to the negative side of  
the circuit.  
Touch the red test probe tip to the positive side of  
the circuit.  
4. Read the voltage in the display. Reset the function  
switch to successively lower V AC positions to obtain  
a higher resolution reading. The display will indicate  
the proper decimal point and value.  
11  
 
DC CURRENT MEASUREMENTS  
CAUTION: Do not make current measurements on the 10A scale for  
longer than 30 seconds. Exceeding 30 seconds may cause damage to  
the meter and/or the test leads.  
1. Insert the black test lead banana plug into the negative  
(COM) jack.  
2. For current measurements up to 200mA DC, set the  
function switch to the highest DC mA position and  
insert the red test lead banana plug into the (mA) jack.  
3. For current measurements up to 10A DC, set the  
function switch to the 10A range and insert the red test  
lead banana plug into the (10A) jack.  
4. Remove power from the circuit under test, then open  
up the circuit at the point where you wish to measure  
current.  
5. Touch the black test probe tip to the negative side of the circuit.  
Touch the red test probe tip to the positive side of the circuit.  
6. Apply power to the circuit.  
7. Read the current in the display. For mA DC measurements, reset the  
function switch to successively lower mA DC positions to obtain a  
higher resolution reading. The display will indicate the proper decimal  
point and value.  
12  
 
RESISTANCE MEASUREMENTS  
WARNING: To avoid electric shock, disconnect power to the unit under  
test and discharge all capacitors before taking any resistance  
measurements. Remove the batteries and unplug the line cords.  
1. Set the function switch to the highest ꢃꢂ position.  
2. Insert the black test lead banana plug into the  
negative (COM) jack  
Insert the red test lead banana plug into the  
positive jack.  
3. Touch the test probe tips across the circuit or  
part under test. It is best to disconnect one side  
of the part under test so the rest of the circuit will  
not interfere with the resistance reading.  
4. Read the resistance in the display and then set  
the function switch to the lowest position that is  
greater than the actual or any anticipated resistance. The display will  
indicate the proper decimal point and value.  
CONTINUITY CHECK  
WARNING: To avoid electric shock, never measure continuity on  
circuits or wires that have voltage on them.  
1. Set the function switch to the  
position.  
2. Insert the black lead banana plug into the  
negative (COM) jack  
Insert the red test lead banana plug into the  
positive () jack.  
3. Touch the test probe tips to the circuit or wire  
you wish to check.  
4. If the resistance is less than approximately 30,  
the audible signal will sound. If the circuit is  
open, the display will indicate “1”.  
13  
 
DIODE TEST  
1. Insert the black test lead banana plug into the  
negative COM jack and the red test lead banana  
plug into the positive diode jack.  
2. Turn the rotary switch to the  
/ •))) position.  
3. Touch the test probes to the diode under test.  
Forward voltage will indicate 400 to 700mV.  
Reverse voltage will indicate “I”. Shorted devices  
will indicate near 0mV. Shorted devices will  
indicate near 0mV and an open device will  
indicate “I” in both polarities.  
BATTERY TEST  
1. Insert the black test lead banana plug into the negative COM jack and  
the red test lead banana plug into the positive V jack.  
2. Select the 1.5V or 9V BAT position using the  
function select switch.  
3. Connect the red test lead to the positive side of  
the 1.5V or 9V battery and the black test lead to  
the negative side of the 1.5V or 9V battery.  
4. Read the voltage in the display.  
Good  
>8.2V  
Weak  
7.2 to 8.2V  
Bad  
<7.2V  
9V battery:  
1.5V battery: >1.35V 1.22 to 1.35V <1.22V  
14  
 
MAINTENANCE  
WARNING: To avoid electric shock, disconnect the test leads from any  
source of voltage before removing the back cover or the battery or fuse  
doors.  
WARNING: To avoid electric shock, do not operate your meter until the  
battery and fuse doors are in place and fastened securely.  
This MultiMeter is designed to provide years of dependable service, if the  
following care instructions are performed:  
1. KEEP THE METER DRY. If it gets wet, wipe it off.  
2. USE AND STORE THE METER IN NORMAL TEMPERATURES.  
Temperature extremes can shorten the life of the electronic parts and  
distort or melt plastic parts.  
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can  
damage the electronic parts or the case.  
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp  
cloth. DO NOT use chemicals, cleaning solvents, or detergents.  
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE  
AND TYPE. Remove old or weak batteries so they do not leak and  
damage the unit.  
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME,  
the batteries should be removed to prevent damage to the unit.  
15  
 
REPLACING THE BATTERIES  
WARNING: To avoid electric shock, disconnect the test leads from any  
source of voltage before removing the battery door.  
1. When the batteries become exhausted or drop below the operating  
voltage, “BAT” will appear in the right-hand side of the LCD display.  
The batteries should be replaced.  
2. Follow instructions for installing batteries. See the Battery Installation  
section of this manual.  
3. Dispose of the old batteries properly.  
WARNING: To avoid electric shock, do not operate your meter until the  
battery door is in place and fastened securely.  
REPLACING THE FUSES  
WARNING: To avoid electric shock, disconnect the test leads from any  
source of voltage before removing the fuse door.  
1. Disconnect the test leads from the meter and any item under test.  
2. Remove the protective rubber holster.  
3. Open the fuse door by loosening the screw on the door using a Phillips  
head screwdriver.  
4. Remove the old fuse from its holder by gently pulling it out.  
5. Install the new fuse into the holder.  
6. Always use a fuse of the proper size and value (0.2A/250V fast blow  
for the 200mA range, 10A/250V fast blow for the 10A range).  
7. Put the fuse door back in place. Insert the screw and tighten it securely.  
WARNING: To avoid electric shock, do not operate your meter until the  
fuse door is in place and fastened securely.  
UL LISTED  
The UL mark does not indicate that this product has been evaluated for the  
accuracy of its readings.  
16  
 
TROUBLESHOOTING  
There may be times when your meter does not operate properly. Here are  
some common problems that you may have and some easy solutions to  
them.  
Meter Does Not Operate:  
1. Always read all the instructions in this manual before use.  
2. Check to be sure the batteries are properly installed.  
3. Check to be sure the batteries are good.  
4. If the battery is good and the meter still does not operate, check to be  
sure that both ends of the fuse are properly installed.  
If You Do Not Understand How the Meter Works:  
1. Purchase the instructional book “Multitesters and Their Use for Electrical  
Testing” (Item No. 82303) at your local Sears store.  
2. Call our Customer Service Line 1-888-326-1006.  
SERVICE AND PARTS  
Item Number  
82374  
Description  
Fuse kit  
93894  
9V battery  
82378  
Set of black and red Test Leads  
Replacement battery door  
Replacement fuse door  
Rear cover screws  
82140-DB  
82140-DF  
82140-CS  
For replacement parts shipped directly to your home  
Call 9 am – 5 pm Eastern Time, M - F  
1-888-326-1006  
17  
 
Owner's Manual  
Voltage Detector Model No. 82174  
CAUTION: Read, understand and follow Safety Rules and Operating  
Instructions in this manual before using this product.  
© Sears, Roebuck and Co., Hoffman Estates, IL 60179 U.S.A.  
022205  
FIVE YEAR FULL WARRANTY  
FIVE YEAR FULL WARRANTY ON CRAFTSMAN Voltage Detector.  
If this CRAFTSMAN Voltage Detector fails to give complete  
satisfaction within five years from the date of purchase, RETURN IT  
TO THE NEAREST SEARS STORE OR OTHER CRAFTSMAN  
OUTLET IN THE UNITED STATES, and Sears will replace it, free of  
charge.  
If this CRAFTSMAN Voltage Detector is used for commercial or rental  
purposes, this warranty applies for 90 days from the date of  
purchase.  
This warranty gives you specific legal rights, and you may also have  
other rights which vary from state to state  
Sears, Roebuck and Co., Dept. 817WA, Hoffman Estates, IL 60179  
For Customer Assistance Call 9am-5 PM (EST), Monday through  
Friday 1-888-326-1006  
18  
 
WARNING: USE EXTREME CAUTION IN THE USE OF THIS  
DEVICE. Improper use of this device can result in injury or death.  
Follow all safeguards suggested in this manual in addition to the  
normal safety precautions used in working with electrical circuits. DO  
NOT service this device if you are not qualified to do so.  
SAFETY SYMBOLS  
This symbol adjacent to another symbol, terminal or  
operating device indicates that the operator must refer to  
an explanation in the Operating Instructions to avoid  
personal injury or damage to the Voltage Detector.  
This WARNING symbol indicates a potentially hazardous  
situation, which if not avoided, could result in death or  
serious injury.  
WARNING  
CAUTION  
This CAUTION symbol indicates a potentially hazardous  
situation, which if not avoided, may result in damage to the  
Voltage Detector.  
This symbol indicates that a device is protected  
throughout by double insulation or reinforced insulation.  
UL LISTED  
The UL mark does not indicate that this product has been evaluated for the  
accuracy of its readings.  
19  
 
SPECIFICATIONS  
Voltage Sensitivity  
Frequency  
100 to 240V AC  
50/60Hz  
Detection distance  
Over voltage  
<0.2”  
Category III 600V  
Operating Temperature 32 to 122oF (0 to 50oC)  
Storage Temperature  
Humidity  
-4 to 140oF (-20 to 60oC)  
<85%RH  
Altitude  
<2000 meters (6561 feet)  
2 AAA batteries  
Battery  
Dimensions/Weight  
Safety  
6x0.75x1 inches / 2oz  
For indoor use and in accordance with  
Overvoltage Category III, Pollution Degree 2.  
BATTERY INSTALLATION  
6. Open the battery door (end cap) by gently  
prying up/out at the pocket clip using a small  
screwdriver.  
7. Insert two AAA batteries (observe polarity).  
8. Replace the battery door.  
End cap  
Removal  
for battery  
replacement  
NOTE: If your meter does not work properly,  
check the battery to make sure that it still  
good and that it is properly inserted.  
20  
 
WARNING: To avoid electric shock, do not operate the meter until the  
battery cover is in place and fastened securely.  
OPERATING INSTRUCTIONS  
WARNING: Risk of Electrocution. High-voltage circuits, both AC and  
DC, are very dangerous and should be measured with great care.  
AC VOLTAGE DETECTION  
WARNING: Risk of Electrocution. Before using to check for voltage in  
an outlet, always test the Voltage Detector on a known live circuit to  
verify that the Voltage Detector is working properly.  
WARNING: Risk of Electrocution. Keep hands and fingers on the body  
of the probe and away from the probe tip  
To check for the presence of AC electrical voltage in an outlet:  
1. Touch the probe tip to a cord plugged into the outlet, or insert the probe  
tip into all the outlet holes.  
2. If AC electrical voltage is present, the detector light will flash and the  
audible warning will sound.  
NOTE: Random beeps and flashes due to detection of static charges are  
normal. Detection of “live” conductors will produce a steady flash  
and beep.  
21  
 

Continental Electric Fan KU33474 User Manual
Cornelius Thermostat WCC1400 User Manual
Craftsman Lawn Mower 107289920 User Manual
Cuisinart Blender SPB 600 User Manual
Cuisinart Coffeemaker PG 30765 User Manual
Curtis Water Dispenser MWC519 User Manual
Dell Server EMC AX4 5I User Manual
Fortinet Network Card FortiMail 2000A User Manual
Frigidaire Dishwasher 685000800012 User Manual
GE Oven JKP75BMBB User Manual